Digital Calibration Certificate

PTX-2025-A5ER-W9RC

Recalibration Due

Eppendorf Research plus, 100–1000 µL

Air-displacement, variable volume

Serial number
EP-5561203
Channels
1
Volume range
100–1000 µL
Registered lab
Harlow Immunology Core
Location
Chicago, IL
Overall result
Pass
Scan to verify
Performed
August 8, 2025
Next due
August 8, 2026
Technician
R. Chen, Senior Metrologist
Standard
ISO 8655-2 / ISO 8655-6
Method
Gravimetric method per ISO 8655-6, 10 replicates per test volume
Reference balance
Sartorius Cubis II MSA, 0.01 mg readability
Conditions
21°C · 45% RH · 1012 hPa

ISO 8655 measurement results

Gravimetric test at 10%, 50%, and 100% of nominal volume, per ISO 8655-6.

Test pointMean measuredSystematic errorRandom error (CV)Result
10% – 100 µL101.50 µL+1.50% / ±3%1.10% / ≤2% Pass
50% – 550 µL552.90 µL+0.53% / ±1.20%0.45% / ≤0.80% Pass
100% – 1000 µL1006.40 µL+0.64% / ±0.80%0.28% / ≤0.50% Pass

Raw measurement data

volumes derived as mass × Z, Z = 1.0031 µL/mg
10% – 100 µL10 weighings
#MASS (mg)VOLUME (µL)
1100.2550100.5658
2100.3021100.6130
3101.4074101.7218
4103.0177103.3371
5101.0674101.3807
699.634199.9430
7101.7477102.0631
8100.4379100.7493
9102.8586103.1775
10101.1353101.4488
101.5000
50% – 550 µL10 weighings
#MASS (mg)VOLUME (µL)
1553.0662554.7807
2547.7793549.4774
3552.4098554.1223
4551.6468553.3569
5554.9483556.6686
6548.8036550.5049
7553.4900555.2058
8552.2019553.9137
9548.1494549.8487
10549.4178551.1210
552.9000
100% – 1000 µL10 weighings
#MASS (mg)VOLUME (µL)
11004.25911007.3723
21005.67561008.7932
31000.55791003.6596
41001.86011004.9659
51003.98951007.1019
61005.24141008.3576
7999.99011003.0901
81000.63601003.7380
91001.86441004.9702
101008.82381011.9512
1006.4000

Calibration history

Every past calibration for this instrument is its own permanent certificate — click any entry to see its full measurement data.

  1. August 8, 2026Calibrated — pass

    M. Alvarez, Lead Metrologist — No adjustment required.

    PTX-2026-A5ER-6DS9

  2. August 8, 2025Calibrated — passViewing now

    R. Chen, Senior Metrologist

    PTX-2025-A5ER-W9RC