Digital Calibration Certificate

PTX-2026-A5ER-6DS9

Valid

Eppendorf Research plus, 100–1000 µL

Air-displacement, variable volume

Serial number
EP-5561203
Channels
1
Volume range
100–1000 µL
Registered lab
Harlow Immunology Core
Location
Chicago, IL
Overall result
Pass
Scan to verify
Performed
August 8, 2026
Next due
August 8, 2027
Technician
M. Alvarez, Lead Metrologist
Standard
ISO 8655-2 / ISO 8655-6
Method
Gravimetric method per ISO 8655-6, 10 replicates per test volume
Reference balance
Sartorius Cubis II MSA, 0.01 mg readability
Conditions
21°C · 45% RH · 1012 hPa

Technician notes: No adjustment required.

ISO 8655 measurement results

Gravimetric test at 10%, 50%, and 100% of nominal volume, per ISO 8655-6.

Test pointMean measuredSystematic errorRandom error (CV)Result
10% – 100 µL101.30 µL+1.30% / ±3%1% / ≤2% Pass
50% – 550 µL552.60 µL+0.47% / ±1.20%0.40% / ≤0.80% Pass
100% – 1000 µL1006 µL+0.60% / ±0.80%0.25% / ≤0.50% Pass

Raw measurement data

volumes derived as mass × Z, Z = 1.0031 µL/mg
10% – 100 µL10 weighings
#MASS (mg)VOLUME (µL)
1103.2760103.5962
2101.2809101.5949
3100.2694100.5802
4101.8446102.1603
599.6981100.0072
6100.1143100.4247
7100.6217100.9336
8100.8092101.1217
9101.0614101.3747
10100.8937101.2065
101.3000
50% – 550 µL10 weighings
#MASS (mg)VOLUME (µL)
1549.0252550.7272
2553.1901554.9050
3549.1031550.8053
4552.3016554.0137
5551.9963553.7075
6547.2891548.9857
7554.0476555.7651
8549.6667551.3707
9549.7451551.4493
10552.5574554.2703
552.6000
100% – 1000 µL10 weighings
#MASS (mg)VOLUME (µL)
11003.40651006.5171
21004.04471007.1572
31007.58151010.7050
41002.14131005.2479
5999.39131002.4894
61005.29531008.4117
71002.65851005.7667
8999.71701002.8161
91003.68731006.7987
101000.98701004.0901
1006.0000

Calibration history

Every past calibration for this instrument is its own permanent certificate — click any entry to see its full measurement data.

  1. August 8, 2026Calibrated — passViewing now

    M. Alvarez, Lead Metrologist — No adjustment required.

    PTX-2026-A5ER-6DS9

  2. August 8, 2025Calibrated — pass

    R. Chen, Senior Metrologist

    PTX-2025-A5ER-W9RC