Digital Calibration Certificate

PTX-2025-2NQY-CF1J

Recalibration Due

Gilson PIPETMAN L P200L

Air-displacement, variable volume

Serial number
GL-4471029
Channels
1
Volume range
20–200 µL
Registered lab
Meridian Diagnostics Lab
Location
Austin, TX
Overall result
Pass
Scan to verify
Performed
September 27, 2025
Next due
August 31, 2026
Technician
M. Alvarez, Lead Metrologist
Standard
ISO 8655-2 / ISO 8655-6
Method
Gravimetric method per ISO 8655-6, 10 replicates per test volume
Reference balance
Sartorius Cubis II MSA, 0.01 mg readability
Conditions
21.2°C · 44% RH · 1016 hPa

Technician notes: Tip cone inspected — no wear detected.

ISO 8655 measurement results

Gravimetric test at 10%, 50%, and 100% of nominal volume, per ISO 8655-6.

Test pointMean measuredSystematic errorRandom error (CV)Result
10% – 20 µL20.32 µL+1.60% / ±2.50%1% / ≤1.50% Pass
50% – 100 µL100.55 µL+0.55% / ±1.20%0.40% / ≤0.70% Pass
100% – 200 µL200.80 µL+0.40% / ±1%0.25% / ≤0.50% Pass

Raw measurement data

volumes derived as mass × Z, Z = 1.00314 µL/mg
10% – 20 µL10 weighings
#MASS (mg)VOLUME (µL)
120.159620.2229
219.982320.0450
320.189320.2527
420.361620.4255
520.116020.1792
620.127320.1905
720.175020.2383
820.652020.7168
920.281720.3454
1020.519220.5836
20.3200
50% – 100 µL10 weighings
#MASS (mg)VOLUME (µL)
1100.6498100.9658
2100.2293100.5440
3100.1625100.4770
499.9999100.3139
5100.1299100.4443
6100.1534100.4679
7100.5974100.9133
8100.1836100.4982
9100.8437101.1603
1099.403299.7153
100.5500
100% – 200 µL10 weighings
#MASS (mg)VOLUME (µL)
1199.8350200.4625
2199.7814200.4087
3200.8585201.4892
4199.5015200.1279
5200.5071201.1367
6200.4726201.1021
7200.7202201.3505
8200.4105201.0398
9199.4601200.0864
10200.1677200.7962
200.8000

Calibration history

Every past calibration for this instrument is its own permanent certificate — click any entry to see its full measurement data.

  1. September 27, 2025Calibrated — passViewing now

    M. Alvarez, Lead Metrologist — Tip cone inspected — no wear detected.

    PTX-2025-2NQY-CF1J

  2. September 27, 2024Calibrated — pass

    R. Chen, Senior Metrologist

    PTX-2024-2NQY-E47V