Digital Calibration Certificate

PTX-2024-2NQY-E47V

Recalibration Due

Gilson PIPETMAN L P200L

Air-displacement, variable volume

Serial number
GL-4471029
Channels
1
Volume range
20–200 µL
Registered lab
Meridian Diagnostics Lab
Location
Austin, TX
Overall result
Pass
Scan to verify
Performed
September 27, 2024
Next due
September 27, 2025
Technician
R. Chen, Senior Metrologist
Standard
ISO 8655-2 / ISO 8655-6
Method
Gravimetric method per ISO 8655-6, 10 replicates per test volume
Reference balance
Sartorius Cubis II MSA, 0.01 mg readability
Conditions
21.2°C · 44% RH · 1016 hPa

ISO 8655 measurement results

Gravimetric test at 10%, 50%, and 100% of nominal volume, per ISO 8655-6.

Test pointMean measuredSystematic errorRandom error (CV)Result
10% – 20 µL20.35 µL+1.75% / ±2.50%1.10% / ≤1.50% Pass
50% – 100 µL100.60 µL+0.60% / ±1.20%0.45% / ≤0.70% Pass
100% – 200 µL200.90 µL+0.45% / ±1%0.28% / ≤0.50% Pass

Raw measurement data

volumes derived as mass × Z, Z = 1.00314 µL/mg
10% – 20 µL10 weighings
#MASS (mg)VOLUME (µL)
120.120620.1838
220.082420.1455
320.059420.1224
420.454720.5189
520.383620.4476
620.192920.2563
720.500620.5650
820.172820.2361
920.739520.8046
1020.156420.2197
20.3500
50% – 100 µL10 weighings
#MASS (mg)VOLUME (µL)
1100.9061101.2229
2100.6930101.0092
399.556699.8692
4100.5386100.8543
5100.7714101.0878
6100.0492100.3634
799.9278100.2416
8100.0208100.3349
999.8941100.2078
10100.4934100.8089
100.6000
100% – 200 µL10 weighings
#MASS (mg)VOLUME (µL)
1200.6069201.2368
2199.4265200.0527
3200.1741200.8026
4199.9737200.6016
5200.2880200.9169
6199.6260200.2528
7201.3996202.0320
8200.0928200.7211
9200.4742201.1037
10200.6496201.2796
200.9000

Calibration history

Every past calibration for this instrument is its own permanent certificate — click any entry to see its full measurement data.

  1. September 27, 2025Calibrated — pass

    M. Alvarez, Lead Metrologist — Tip cone inspected — no wear detected.

    PTX-2025-2NQY-CF1J

  2. September 27, 2024Calibrated — passViewing now

    R. Chen, Senior Metrologist

    PTX-2024-2NQY-E47V